2. Press to mark this test as a control solu-
tion test
With “QC” displays, the meter will not store your
test result in memory. If you press
again, the
“QC” will disappear and this test is no longer a
control solution test.
3. Apply Control Solution
Shake the control solution vial thoroughly before
use.
Squeeze out a drop and wipe it off, then squeeze
another drop and place it on the tip of the vial cap.
Hold the meter to move the absorbent hole of test strip to touch the
drop. Once the conrmation window lls completely, the meter will
begin counting down.
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.
When doing the control solution test, you have to mark it so that
the test result will not be stored in the memory. Failure to do so
will mix up the blood glucose test results with the control solution
test results in memory.